PLANET 55 PORTABLE MONITOR TECHNICAL SPECIFICATIONS
DISPLAY TYPE: High resolution Color TFT Display SIZE: 8.4 (diagonally) RESOLUTION: 800 X 600 pixels excellent viewing from distance and angle PARAMETERS: ECG, respiration, SPO2 DUAL NIBP, temperature& CO2 TRACE SPEED: 12.5, 25 & 50 MM/SEC. for ECG & SPO2 6.25, 12.5 & 25 MM/SEC. for respiration & CO2
OPTICAL ENCODER / KEYS OPTICAL ENCODER: rapid access to all the functions and settings of parameters through single knob HOT KEYS: quick action hot keys for alarm acknowledge, nibp start/stop, recorder start/stop, freeze, stand by, go-to and return to main screen
ECG MONITORING LEAD: 3/5 lead with optional st & arrhythmia
PROTECTION: built-in cautery & defibrillator protection HR RANGE: 20 to 350 BPM ( +2 BPM or 2% whichever is greater)
RESPIRATION (FROM ECG) PRINCIPLE: impedance pneumography RR RANGE: 4 to150 BPM
PULSE OXIMETRY (NELLCOR MODULE) TONE VARIATION: tone variation with change in spo2 values: 1-100% ACCURACY: adult 100-70% ( +2 digits) 0-69% (unspecified) NEONATES: 100-70% ( +3 digits) PERFUSION INDICATOR: bar graph showing signal strength
MODES: normal response & fast response ALARMS: low 0-95% high5-100% PR RANGE: 20 to 230 BPM
TEMPERATURE CHANNEL: dual channel RANGE: 0°C to50°C UNIT: °C or°F probe compatible with ysi 400 series probes (optional)
NON-INVASIVE BLOOD PRESSURE (SUNTECH MODULE) PRINCIPLE: oscillometric DISPLAY: systolic, diastolic & mean pressure
MODES: manual, stat (continuous 5 min. operation) and automatic (time interval 2-90 min. selectable) ALARM LIMITS: selection possible for systolic & diastolic RANGE: 20-250 MMHG ACCURACY: +3 MMHG
CAPNOGRAPHY (ORIDION MODULE) microstream capnography with display of CO2 waveform and digital values ofEtCO2, FiCO2 & RR CO2 gets priority for calculating respiration rate EtCO2 RANGE: 0-99 MMHG FiCO2 RANGE: 0-20 MMHG GAIN: 20,40,60,80 MMHG FLOWRATE: 50 ML/MIN (-7.5 /-15 ML/MIN) WARM-UP TIME: 40 SEC. (typically), units: MMHG /KPA / VOL% RR RANGE: 0-150 BPM
ALARMS ALARM SETTING FOR ALL PARAMETERS PATIENT ALARM: red flashing with audio and message indicator
EQUIPMENT ALARM: yellow flashing with audio and message indicator ALARM SUSPEND: continuous yellow
AUTO SETTING: HR / PR, SPO2, RR, temperature & CO2 MANUAL SETTING: HR / PR, SPO2 systolic, diastolic, RR. temperature & CO2 RE-ALARM: 3 min. for acknowledge alarm
TRENDS 72 HRS / 100 patients graphical and tabular trend for HR / PR, SPO2, RR temperature, CO2 & ST NIBP TREND: last 240 readings (tabular) ALARM RECALL: last 12 critical alarms with time & date
RECORDING (OPTIONAL) TYPE: built-in channel thermal array recorder for recording ecg, plethysmograph, respiration & co2 waveforms (any two-uses selectable) with numerical data, nibp & tabular trend recording MODES: direct / delayed / continuous and record on alarm features
GENERAL SLAVE DISPLAY DIMENSION: possible with external monitor (H X W X D): 235 X 260 X 158 MM WEIGHT: 4.5 KG(with built-in battery & thermal array recorder)
POWER SUPPLY UNIVERSAL POWER: 95-264V AC, 10-165 HZ ( +5%) CURRENT RATING: 0.7A (MAX) BATTERY TYPE: 14.8V (4AH /8AH) Lithium Ion (Built-in) BATTERY OPTIONS: rechargeable battery for 2/4 hrs of continuous operation under full charge condition (typically)
CHARGE TIME: 8 HRS (typically)
|