LAF386390061027
L&T Star 55 Color Patient Monitor
FIVE YEAR PARTS AND LABOR WARRANTY
DISPLAY
TYPE: HIGH RESOLUTION COLOR TFT DISPLAY
RESOLUTION: 800 X 600 PIXELS
SIZE: CONFIGURABLE SCREEN FOR MAX OF EIGHT WAVEFORMS
TRACE SPEED: 12.5, 20 OR 50 MM/SEC FOR ECG, SpO2 & IBP 6.25, 12.5 OR 25 MM/SEC FOR RESPIRATION & CO2
OPTICAL ENCODER KEYS
OPTICAL ENCODER: RAPID ACCESS TO ALL THE FUNCTIONS AND SETTINGS THROUGH SINGLE KNOB
HOT KEYS: QUICK ACTION HOT JEYS FOR ALARM ACKNOWLEDGE, NIBP START/STOP, RECORD START/STOP, IBP ZERO, FREEZE, STAND BY, GO-TO, CO2 PUMP OFF AND HOME KEY
3/5 LEAD ECG MONITORING
AUTOMATIC DETECTION OF 3 OR 5 LEAD ECG CABLE. DISPLAY OF I, II, III WITH 3 - LEAD CABLE AND I, II, III AVR, AVL, AVF AND V LEADS WITH 5 - LEAD CABLE. BUILT-IN CAUTERY AND DEFIBRILLATOR PROTECTION. OPTIONAL ST AND ARRHYTHMIA ANALYSIS
HR RANGE: 20 TO 350 BPM
RESPIRATION
PRINCIPLE: IMPEDANCE PNEUMOGRAPHY. DISPLAY OF RESPIRATION WAVEFORM WITH RESPIRATION RATE
RR RANGE: 4 TO 150 BPM
TEMPERATURE
SIMULTANEOUS MONITIRING OF TWO TEMPERATURES. DISPLAY OF T1 AND T2 OR ΔT (IF DIFFERENCE BETWEEN T1 & T2 IS LESS THAN 9.9°C)
PROBE: COMPATIBLE WITH YSI 400 SERIES TEMP. PROBES (OPTIONAL)
TEMP. RANGE: 0°C TO 43°C
PULSE OXIMETRY (NELLCOR MODULE)
PITCH VARIATION: WITH CHANGE IN OXYGEN SATURATION
SpO2 RANGE: 0-100%
DISPLAY: PLETHYSMOGRAPH, PULSE STRENGTH & SpO2 VALUES.
ACCURACY: ADULT - 100 TO 70% (+/- 2 DIGITS), 0 TO 69% (UNSPECIFIED) (+/- STD. DEV.)
NEONATE: 100 TO 70% (+/- 3 DIGITS), 0 TO 69% (UNSPECIFIED)
PR RANGE: 20 TO 230 BPM +/- 3 BPM (+/- STD. DEV.)
INVASIVE BLOOD PRESSURE
SIMULTANEOUS MONITORING OF MAX. 4 CHANNELS. DISPLAY OF SYSTOLIC, DIASTOLIC AND MEAN PRESSURE. INDIVIDUAL AND COMMON CHENNEL ZEROING. LABELING FACILITY WITH AUTOMATIC LIMITS/GAIN/COLOR SETTING. PRESSURE CHANNEL OVERLAPPING OPTION AVAILABLE
RANGE: -40 TO 300 MMHG
IBP SCALE: 5, 10, 15, 20, 30, 60, 80, 100, 120, 150, 180, 200, 240, 300 & AUTO
NON-INVASIVE BLOOD PRESSURE (SUNTECH MODULE)
PRINCIPLE: OSCILLOMETRIC
DISPLAY: SYSTOLIC, DIASTOLIC & MEAN PRESSURE
MODES: MANUAL, STAT (5 MIN), AUTO. (2 MIN TO 90 MIN)
RANGE: 20 TO 250 MMHG
ACCURACY: +/- 3 MMHG
CAPNOGRAPHY (ORIDION MODULE)
MICROSTREAM CAPNOGRAPHY WITH DISPLAY OF CO2 WAVEFORM AND DIGITAL VALUES OF EtCO2, FiCO2, & RR, CO2 GETS PRIORITY FOR CALCULATING RESP. RATE
EtCO2 RANGE: 0 - 99 MMHG
FiCO2 RANGE: 0 - 20 MMHG
GAIN: 20, 40, 60, 80 MMHG
FLOW RATE: 50 ML/MIN (-7.51 / + 15ML/MIN)
WARM-UP TIME: 40 SECS (TYPICALLY)
UNITS: MMHG, KPA / VOL%, RR
RANGE: 0 -150 BPM
ALARMS
ALARM SETTINGS FOR ALL PARAMETERS
PATIENT ALARM: RED FLASHING WITH AUDIO AND MESSAGE INDICATOR ALARM AUTO-SET FACILITY PROVIDED
EQUIPMENT ALARM: YELLOW FLASHING WITH AUDIO AND MESSAGE INDICATOR
ALARM SUSPEND: CONTINUOUS YELLOW
RE-ALARM FACILITY: 3 MIN FOR ACKNOWLEDGED ALARM
TRENDS
72 HRS/100 PATIENTS GRAPHICAL AND TABULAR TREND WITH ZOOM FACILITY
NIBP TREND: 240 READINGS TABULAR
ALARM RECALL: LAST 24 ALARM EVENTS WITH DATE, TIME & MESSAGE
RECORDER (OPTIONAL)
TYPE: BUILT-IN CHANNEL THERAML RECORDER FOR RECORDING TWO WAVEFORMS (SELECTABLE) WITH NUMERICAL DATA AND ALARM STATUS. NIBP AND TABULAR TREND RECORDING WITH PATIENT INFORMATION
MODES: DIRECT / DELAYED / CONTINUOUS AND RECORD ON ALARM FACILITY (SELECTABLE)
GENERAL
DIMENSION: (H X W X D) 208 X 320 X 158 MM
WEIGHT: 6KG (WITH BUILT-IN BATTERY & THERMAL ARRAY RECORDER)
POWER SUPPLY
UNIVERSAL POWER: 95 - 265V AC, 50/60 Hz
CURRENT RATING: 0.7 A (MAX)
BATTERY TYPE: 14.8V (4 AH/8 AH) LITHIUM ION BATTERIES
OPERATING TIME: BUILT-IN RECHARGABLE BATTERIES FOR 2/4 HRS OF CONTINUOUS OPERATION UNDER FULL CHARGE CONDIDTION (TYPICALLY)
CHARGING TIME: 8 HRS (TYPICALLY)
|