Display:
High resolution color TFT: size (6.5" diagonal)
Resolution: (640 x 480 dots)
Excellent Viewing from distance and angle.
Sharp and high contrast TFT display technology.
Wave form field:
1st trace ECG, 2nd trace cascade ECG/pleth. and 3rd trace plethysmograph (configurable).
Digital field:
HR/PR, SpO2, NIBP (systolic, Diastolic & Mean) and Temperature. On-screen display of alarm limits or alarm disable status.
Trace speed:
12.5, 25 or 50 mm/sec. for both ECG and Plethysmoprapgh.
Optical Encoder/Keys:
Rapid access to all the functions and setting of parameters through single-knob. Quick action hot-keys for alarm recall, NIBP start/stop, recorder start/stop, freeze and return to the main screen. Smart Alarm acknowledge key with flash indicator.
ECG Monitoring:
3 lead ECG monitoring (I,II,III) with cascade wave form facility.
Built-in cautery and defibrillator protection.
HR range: 20 to 270 bpm
Bandwidth (monitoring): 0.5 Hz - 25 Hz
Bandwidth (monitoring): 0.05 Hz - 25 Hz
Pulse Oximetry:
Pitch variation with change in oxygen saturation.
SpO2 range: 0 - 100%
Display of Plethysmograph, pulse strength and SpO2 values.
Accuracy:
100 - 70% (+/- 2 digits),
69 - 50% (+/- 3 digits),
49 - 0% (unspecified).
Accuracy in neonatal monitoring 100 - 70% (+/- 3 digits).
Non-Invasive Blood Pressure:
Oscillometric principle. Display of Systolic, Diastolic & Mean.
Modes: Manual, STAT (continuous 5-minute operation) and Automatic (time interval 2-minutes to 90 minute selectable).
Alarm Limits selection possible for Systolic & Diastolic.
Range: 20 to 255 mmHg
Accuracy: +/- 5 mmHg
Temperature:
(compatible with YSI 400 series Temp. probes)
Single temperature monitoring (ºC or ºF)
Temperature range: 53.6 ºF to 109.4 ºF (12 ºC to 43 ºC)
Alarms:
Alarm setting for all parameters.
Patient alarm: RED Flashing with audio and message indicator.
Equipment alarm: YELLOW flashing with audio and message indicator.
Alarm suspend: Continuous YELLOW.
Auto setting of alarms for HR/PR, SpO2, Systolic & Diastolic and Temp.
Re-Alarm facility: 3 minutes for acknowledged alarm.
Trends:
24-hours of graphical and tabular trend for HR/PR and SpO2 and temp.
NIBP trend: 240 reading tabular.
Alarm recall: last 16 alarm events with time, date and message.
Trends: 30 second resolution. 2 hour zoom for graphical trends with mark and print facility.
Recording/Printing:
Built-in, dual-channel thermal array recorder. Optional accessory for recording ECG & pleth wave forms with numerical data. Direct/delayed/ continuous and record-on alarm features. Printing of ECG strip along with complete numerical data and patient information on ink-jet printer possible (optional accessory).
General:
Facility to interface with external monitor.
Dimension: 7.7" H x 9.6" W x 9.8" D (195 x 245 x 250 mm)
Weight: 11 lbs (5 Kg) with built-in battery and thermal array recorder
Direct interfacing facility with "Skyline."
Power Supply:
Universal power: 95V-265V AC, 50/60 Hz
Power ratings: 40W
Built-in, rechargeable battery for 45 minutes of continuous operation under ideal & full-charge condition (typical).
Battery type: 12V, 1.2 Ah sealed lead acid - No. 1
Charging time: 80% charging in 4-hr. (typical)
Ambulatory monitoring possible with car battery interface.
Designed to meet the requirements of Following international standards:
Class 1, Type CF of IEC 601-1
Level 2 of IEC 1000-4-4 and IEC 10004-5