LAF390390006571
L&T Galaxy Color Anesthesia Medical Monitor
FIVE YEAR PARTS AND LABOR WARRANTY
Display:
(640 x 480 pixel) RGB Compatible with VGA monitor (CRT / TFT LCD)
Wave form field:
Configurable screen for maximum 4 wave forms (ECG, Pleth, Resp, Invasive BP or Capnography). Possible to view any 4 ECG leads, if 5-lead ECG cable is connected.
Digital field:
HR/PR, Resp, SpO2, Temp (T1, T2 or ?T) P1, P2 and NIBP (systolic, Diastolic and Mean), EtCO2 and FiCO2. On screen display of alarm limits or alarm disable status.
Trace speed:
12.5, 25 or 50 mm/sec for ECG, Invasive pressure waveform and Plethysmograph. 6.25, 12.5 or 25 mm/sec for Respiration and side Capnogram (with 14", 640 480 pixel display)
Optical Encoder:
Rapid access to all the functions and parameter settings through single knob. Quick action hot keys for alarm mute, NIBP start/stop, IBP zero, recorder on/off, freeze and return to the main screen.
3/5 lead ECG monitoring:
Automatic detection of 3 or 5 lead ECG cable. Display of I, II, III with 3-lead cable and I, II, III, AVR, aVL, aVF and V with 5-lead cable. In built cautery and defibrillator protection.
HR range 20 to 270 bpm
Programmable arrhythmia detection (Bradycardia, Tachycardia, Asystole, Irregular and skipped beat)
Respiration:
Impedance pneumography principle.
Display of respiration waveform with respiration rate.
Respiration range: 4 to 150 bpm
User selectable aponea alarm delay: 5 - 90 sec. (steps of 5 sec.)
Capnography (CO2):
Side stream Capnography with display of CO2 waveform and digital values of EtCO2 and FiCO2.
CO2 gets priority for calculation of respiration rate.
Gain 20,40,60 and 80 mmHg. Flow rate: 140 ml/min. (+20 / -10)
CO2 measurement is possible in mmHg /Kpa / Vol%
EtCO2 range: 0 - 80 mmHg
FiCO2 range: 2 - 20 mmHg
Aponea alarm delay: 30 sec
Temperature:
Simultaneous monitoring of two temperatures. Display of T1 and T2 or ?T (if difference between T1 and T2 is less than 9.9ºC)
Temperature range: 12 ºC to 43 ºC
Pulse Oximetry:
Pitch variation with change in oxygen saturation.
SpO2 range: 0 - 100%
Display of Pulse strength, plethysmograph and SpO2 values.
Accuracy:
100 to 70% (+/- 2 digits)
69 to 50% (+/- 3 digits)
49 to 0% (unspecified)
Accuracy in neonatal monitoring 100 to 70% (+/- 3 digits)
Non-invasive Blood Pressure:
Oscillometric Principle.
Display of Systolic, Diastolic and Mean pressure.
Modes: manual, STAT (5 min), Automatic (interval 2 min to 90 min)
Range: 20 to 255 mmHg.
Accuracy: +/- 5 mmHg
Invasive Blood Pressure:
Simultaneous monitoring of 2 channels.
Display of Systolic, Diastolic and Mean pressure with respective waveform of P1 and P2. Individual and common channel zeroing. Labeling facility with automatic limits /gain/color setting.
Pressure channel overlapping option available.
IBP range: 300, 150, 60, 30, 20 and 10 mmHg
Measurement range: -40 to 300 mmHg
Alarms:
Alarm setting for all parameters.
Patient alarm: Red flashing with audio and message indicator
Equipment alarm: YELLOW flashing with audio and message indicator. Alarm auto set facility provided.
Alarm suspend: Continuous YELLOW
Auto setting of alarm for HR/PR, SpO2, P1, P2, EtCO2, FiCO2, Respiration and Temperature.
Re-alarm facility: 30 min for acknowledged alarm.
Trends:
24 hr of graphical and tabular trend for HR/PR, SpO2, P1, P2, EtCO2, FiCO2, respiration and temperature.
NIBP trend: 240 reading tabular
Alarm recall: Last 16 alarm events with date, time and message.
2hr zoom (30 sec resolution) for graphical trends with mark, record and print facility.
Recording / Printing:
Optional in-built dual channel thermal recorder for recording two waveforms (selectable) with numerical data and alarm status.
NIBP and tabular trend recording with patient information.
Direct / delayed / Continuous modes of recording.
Record on alarm facility (selectable)
Printing of ECG strip along with complete numerical data and patient information on optical ink-jet printer.
General:
Dimension: 100 H x 350 W x 300 D mm
Weight: 8.3 kg with in-built batteries, Capnography and recorder.
Interface possible with Skyline - L&T color central station.
Power Supply:
Universal power supply: 95-265V ac, 50/60 Hz
Power ratings: 40 W
Operating time: 30 min for fully charged battery (with extreme condition: NIBP 2 min interval, Capnography and recorder in use), or approx 2 hr under full charge condition (typical)
Battery type: 12V, 4Ah lead acid
Charging time: 90% charge in 12 hr for fully discharged battery
*UPS is recommended for providing back up to the external monitor, in case of main power failure. "Galaxy" does not offer battery back up for the external monitor.
*Technological upgradation is a continuous process at L&T. Hence, specifications and availability of any of the products mentioned in this leaflet are subject to change without notice. Some of the accessories featured in this leaflet are optional and are not part of the standard scope of supply.
|