FHC-VALLEYLABFX
FHC Valleylab FX Electrosurgical Generator
OUTPUT WAVEFORMS
BIPOLAR
PRECISE: 470 KHZ SINUSOID
STANDARD: 470 KHZ SINUSOID
MACRO: 470 KHZ SINUSOID
MONOPOLAR CUT
LOW: 390 KHZ SINUSOID. SIMILAR TO THE PURE CUT MODE EXCEPT THE MAXIMUM VOLTAGE IS LIMITED TO A LOWER VALUE.
PURE: 390 KHZ SINUSOID
BLEND: 390 KHZ BURSTS OF SINUSOID, RECURRING AT 27 KHZ INTERVALS. 50% DUTY CYCLE ENVELOPE.
MONOPOLAR COAG
DESICCATE: 240 KHZ SINUSOID REPEATED AT 39 KHZ. 8% DUTY CYCLE.
FULGURATE: 390 KHZ DAMPED SINUSOID BURSTS WITH A RANDOMIZED REPETITION CENTERED AT 28 KHZ. FREQUENCIES INCLUDE 21 KHZ
OUTPUT POWER CHANGES BY LESS THAN 15% OR 5 WATTS, WHICHEVER IS GREATER, AS THE LINE VOLTAGE VARIES FROM 104-132 VOLTS AND 208 - 264 VOLTS (AT RATED LOAD).
LOW FREQUENCY LEAKAGE (50 - 60 HZ)
SOURCE CURRENT, PATIENT LEADS, ALL OUTPUTS TIED TOGETHER.
-
NORMAL POLARITY, INTACT CHASSIS GROUND < 10 µA
-
NORMAL POLARITY, GROUND OPEN < 50 µA
-
REVERSE POLARITY, GROUND OPEN < 50 µA
-
SINK CURRENT, 140V APPLIED, ALL INPUTS < 50 µA
PER = 98
PER (POWER EFFICIENCY RATING) IS THE MEASURE OF AN ELECTROSURGICAL GENERATOR'S ABILITY TO ACCURATELY DELIVER THE SELECTED POWER INTO A WIDE RANGE OF TISSUE TYPES.
WEIGHT AND DIMENSIONS
HEIGHT: 11.1 CM (4 3/8")
WIDTH: 35.6 CM (14")
LENGTH: 43.9 CM (17")
WEIGHT: < 8.1 KG (< 18 LBS)
INPUT POWER REQUIREMENTS
OPERATING RANGE IS 85 TO 132 AC VOLTS. MAXIMUM CURRENT IS 7 AMPERES IN CUT AND 4 AMPERES IN COAG.
HIGH FREQUENCY LEAKAGE
BIPOLAR: LESS THAN 60 M ARMS
MONOPOLAR: LESS THAN 150 M ARMS
REM™ CONTACT QUALITY MONITORING SYSTEM
MEASUREMENT FREQUENCY: 80 KHZ ± 10 KHZ
MEASUREMENT CURRENT: LESS THAN 10 µA
ACCEPTABLE RESISTANCE RANGES:
REM™ PAD - 5-135 OHMS
NON-REM™ PAD - LESS THAN 20 OHMS
ACCEPTANCE RANGE IS 5-135 OHMS AFTER REM POLYHESIVE™ II RETURN ELECTRODE IS APPLIED.
ADAPTIVE REM™
REM™ TRIP IS BASELINE IMPEDANCE PLUS 40%. FOR EXAMPLE, IF THE BASELINE IMPEDANCE IS 30 OHMS, THE UPPER LEVEL TRIP IS APPROXIMATELY 42 OHMS. IF THE PAD-PATIENT IMPEDANCE FALLS BELOW THE BASELINE IMPEDANCE, A NEW BASELINE IS ESTABLISHED
|
MODE |
MAXIMUM P-P VOLTAGE |
RATED LOAD (OHMS) |
MAXIMUM POWER (WATTS) |
CREST FACTOR (TYPICAL) |
|
|
|
|
|
|
| BIPOLAR |
PRECISE |
450 |
100 |
70 |
1.5 |
|
STANDARD |
320 |
100 |
70 |
1.5 |
|
MACRO |
750 |
100 |
70 |
1.5 |
| MONOPOLAR CUT |
LOW |
1350 |
300 |
300 |
1.5 |
|
PURE |
2300 |
300 |
300 |
1.5 |
|
BLEND |
3300 |
300 |
200 |
2.5 |
| MONOPOLAR COAG |
DESICCATE |
3500 |
500 |
120 |
5.0 |
|
FULG. HIGH CRESTFACTOR |
8500 |
500 |
120 |
7.0 |
|
FULG. LOW CREST FACTOR |
6900 |
500 |
120 |
5.5 |
|
SPRAY |
9000 |
500 |
120 |
8.0 |
| CEM MODE |
MONOPOLAR CUR (LOW) |
1000 |
300 |
100 |
1.5 |
|
MONOPOLAR COAG (DESICCATE) |
3500 |
500 |
70 |
5.0 |
*CREST FACTOR IS AN INDICATOR OF A WAVEFORM'S ABILITY TO COAGULATE WITHOUT CUTTING.
|