FHC Valleylab EZ Electrosurgical Generator
OUTPUT WAVE FORMS:
BIPOLAR: 470 KHZ SINUSOID
MONOPOLAR CUT:
PURE: 393 KHZ SINUSOID
BLEND: 393 KHZ BURSTS OF SINUSOID, RECURRING AT 27 KHZ INTERVALS. 50% DUTY CYCLE ENVELOPE
MONOPOLAR COAG:
DESICCATION:
LOW 1: 240± 40 KHZ SINUSOID RECURRING AT 39 KHZ 8% DUTY CYCLE
LOW 2: 393 KHZ SINUSOID
LOW 3: 393 KHZ SINUSOID
FULGURATION:
HIGH 1: 470 ± 40 KHZ DAMPED SINUSOIDAL BURSTS WITH REPETITION FREQUENCY OF 57 KHZ
HIGH 2: 470 ± 40 KHZ DAMPED SINUSOIDAL BURSTS WITH REPETITION FREQUENCY OF 30 KHZ
OUTPUT POWER CHANGES BY LESS THAN 15% OR 5 WATTS, WHICHEVER IS GREATER, AS THE LINE VOLTAGE VARIES FROM 90-135 VOLTS AND 186-264 VOLTS (INTO A 300 OHM LOAD)
LOW FREQUENCY LEAKAGE (50-60 HZ)
SOURCE CURRENT, PATIENT LEADS, ALL OUTPUTS TIED TOGETHER.
-
NORMAL POLARITY, INTACT CHASSIS GROUND < 10 µA
-
NORMAL POLARITY, GROUND OPEN < 50 µA
-
REVERSE POLARITY, GROUND OPEN < 50 µA
-
SINK CURRENT, 140V APPLIED, ALL INPUTS < 50 µA
POWER READOUTS
AGREE WITH ACTUAL POWER INTO RATED LOAD TO WITHIN ± 15% OR 5 WATTS, WHICHEVER IS GREATER. DOSAGE ERROR MONITOR.
PER = 97
PER (POWER EFFICIENCY RATING) IS THE MEASURE OF AN ELECTROSURGICAL GENERATOR'S ABILITY TO ACCURATELY DELIVER THE SELECTED POWER INTO A WIDE RANGE OF TISSUE TYPES.
WEIGHT AND DIMENSIONS
HEIGHT: 12.7 CM (5")
WIDTH: 40.6 CM (16")
LENGTH: 39.5 CM (15.6")
WEIGHT: < 6.8 KG (< 15 LBS)
INPUT POWER REQUIREMENTS
110-120 VAC 220-240 VAC
OPERATING RANGE 58-140 VAC 170-280 VAC
MAX CURRENT CUT 8A 4A
MAX CURRENT COAG 4.2A 2.1A
HIGH FREQUENCY LEAKAGE
BIPOLAR: LESS THAN 60 M ARMS
MONOPOLAR: LESS THAN 150 M ARMS
REM™ CONTACT QUALITY MONITORING SYSTEM
MEASUREMENT FREQUENCY: 80 KHZ ± 10 KHZ
MEASUREMENT CURRENT: LESS THAN 10 µA
ACCEPTABLE RESISTANCE RANGES:
REM™ PAD - 5-135 OHMS
NON-REM™ PAD - LESS THAN 20 OHMS
ACCEPTANCE RANGE IS 5-135 OHMS AFTER REM POLYHESIVE™ II RETURN ELECTRODE IS APPLIED.
ADAPTIVE REM™
REM™ TRIP IS BASELINE IMPEDANCE PLUS 40%. FOR EXAMPLE, IF THE BASELINE IMPEDANCE IS 30 OHMS, THE UPPER LEVEL TRIP IS APPROXIMATELY 42 OHMS. IF THE PAD-PATIENT IMPEDANCE FALLS BELOW THE BASELINE IMPEDANCE, A NEW BASELINE IS ESTABLISHED.
|
MODE |
MAXIMUM P-P VOLTAGE |
RATED LOAD (OHMS) |
MAXIMUM POWER (WATTS) |
CREST FACTOR (TYPICAL) |
|
|
|
|
|
|
| BIPOLAR |
STANDARD |
300 |
100 |
70 |
1.5 |
|
PURE |
2000 |
300 |
300 |
1.5 |
|
BLEND |
3400 |
300 |
200 |
2.1 |
| MONOPOLAR CUT |
DESICCATE LOW 1 |
3500 |
500 |
120 |
5.0 |
|
LOW 2 |
660 |
300 |
120 |
1.5 |
|
LOW 3 |
1100 |
300 |
120 |
1.5 |
|
FULGURATE HIGH 1 |
6200 |
500 |
120 |
5.0 |
|
HIGH 2 |
8500 |
500 |
120 |
7.0 |
*CREST FACTOR IS AN INDICATOR OF A WAVEFORM'S ABILITY TO COAGULATE WITHOUT CUTTING.
|