FHC-EXCALIBUR+
FHC Excalibur Plus Electrosurgical Generator
MAINS OVERCURRENT PROTECTION:
3AG 10A FUSE IN 115V POSITION. 5X20MM 4A QUICK ACTING/HIGH BREAKING FUSE IN 230V POSITION.
MAINS FREQUENCY LEAKAGE:
(PER UL 544) PATIENT CONNECTIONS TO NEUTRAL: LESS THAN 10 µA TYPICAL.
CHASSIS TO NEUTRAL: LESS THAN 25 µA TYPICAL.
POWER DISPLAY ACCURACY:
THE GREATER OF +/- 10% OF DISPLAY OR +/- 3 WATTS AT RATED LOAD.
LINE REGULATION:
POWER CHANGE < 1%/V FOR VLINE = MIN TO MAX VOLTAGE
DUTY CYCLE: 15S ON, 30S IDLE
OPERATING CONDITIONS: 10°C TO 30°C, 95% RH NON-CONDENSING
MOUNTING RESTRICTIONS: 2" (5 CM) CLEARANCE REQUIRED ON EACH SIDE OF UNIT FOR COOLING.
PAD MONITOR
SINGLE PAD MODE: TWO WIRE CONTINUITY DETECTOR, TYPICAL THRESHOLD = 10 OHMS
DUAL PAD MODE: TWO WIRE RESISTANCE MONITOR, TYPICAL ACCEPTANCE RANGE 10 TO 150 OHMS, TRIP THRESHOLD TYPICALLY 20% HIGHER THAN LAST MONITOR SET POINT ACTIVATION, VISUAL INDICATION OF PATIENT RESISTANCE.
SOUND FREQUENCIES:
CUT = 520 HZ
COAG = 400 HZ
OPERATOR ERROR = 1710 HZ, PULSATING
PLATE ALARM = 1000 HZ (65 DBA MINIMUM)
MACHINE FAULT = 1000 HZ (45 DBA MIN EXCEPT ALARMS ARE 65 DBA MIN)
WEIGHT: 40 LB (18.1 KG)
HEIGHT: 6.25" (15.9 CM)
WIDTH: 15.9" (40.4 CM)
DEPTH: 22.75" (57.8 CM)
|
MODE |
MAX POWER (WATTS) |
RATED LOAD (OHMS) |
TYPICAL CREST FACTOR |
MAX OPEN CIRCUIT VOLTAGE (P-P) |
CARRIER REQ. (KHZ) |
PULSE REPETITION FREQ (KHZ) |
|
MONOPOLAR |
|
|
|
|
|
|
|
PURE CUT |
300 |
300 |
1.7 - 1.9 |
2000 |
416.7 |
CONTINUOUS |
|
BLEND 1 CUT |
180 |
300 |
2.3 - 2.5 |
2000 |
416.7 |
20.0 |
|
BLEND 2 CUT |
120 |
300 |
2.9 - 3.2 |
2200 |
416.7 |
20.0 |
|
BLEND 3 CUT |
80 |
300 |
3.6 - 4.0 |
2400 |
416.7 |
20.0 |
|
STD COAG |
120 |
300 |
6.2 - 6.5 |
6500 |
540 |
40.0 |
|
SPARY COAG |
80 |
500 |
8.0 - 9.0 |
10500 |
540 |
20.0 |
|
BIPOLAR |
|
|
|
|
|
|
|
CUT |
50 |
50 |
1.6 - 1.8 |
400 |
1000 |
CONTINUOUS |
|
COAG |
50 |
50 |
1.5 - 12.0 |
400 |
1050 |
20.0 |
|
|
|
|
BIPOLAR COAG CREST FACTOR DECREASES WITH POWER SETTING |
|
|
|
|