FHC Medical EquipmentAbout Future Health ConceptsRequest Medical Equipment Catalog
Free Medical Equipment Catalog FHC Medical Equipment
FHC Medical Equipment

Featured Products
Anesthesia Machines
Blanket Warmers
Fluid Warmer
Crash Carts / Anesthesia Carts
Cart Accessories
Defibrillators
Electrosurgical Units
Medical Gas Systems
Patient Monitors
Scrub Sinks
Stainless Steel Products
Stretchers
Sterilizers / Autoclaves
Surgical Lights
Surgical Tables
Surgical Table Accessories
Ultrasonic Cleaners
X-Ray Supplies and Equipment

Full Medical Equipment List
Replacement Medical Equipment Parts

Electrosurgical Units

Call Future Health Concepts at 1-888-282-8644

FHC Valleylab FX Electrosurgical Generator

FHC Valleylab FX Electrosurgical Generator - Electrosurgical Units
FHC-VALLEYLABFX
FHC Valleylab FX Electrosurgical Generator
OUTPUT WAVEFORMS
   
BIPOLAR
PRECISE: 470 KHZ sinusoid
STANDARD: 470 KHZ sinusoid
MACRO: 470 KHZ sinusoid
MONOPOLAR CUT
 
LOW: 390 KHZ sinusoid. Similar to the pure cut mode except the maximum voltage is limited to a lower value.
PURE: 390 KHZ sinusoid
BLEND: 390 KHZ bursts of sinusoid, recurring at  27 KHZ intervals
 
50% duty cycle envelope.
MONOPOLAR COAG
 
DESICCATE: 240 KHZ sinusoid repeated at 39 KHZ. 8% duty cycle.
FULGURATE: 390 KHZ damped sinusoidal bursts witha randomized repetition centered at 28  KHZ. Frequencies include 21 KHZ.
 
Output power changes by less than 15% or 5 watts, whichever is greater, as the line voltage varies from 104-132 volts and 208 - 264 volts (at rated load).
 
 
 
LOW FREQUENCY LEAKAGE (50 - 60 HZ)
 

Source current, patient leads, all outputs tied together.
Normal polarity, intact chassis ground < 10 µA
Normal polarity, ground open< 50 µA
Reverse polarity, ground open< 50 µA
Sink current, 140V applied, all inputs < 50 µA 
 
 
PER = 98
 

Per (power efficiency rating) is the measure of an electrosurgical generator's ability to accurately deliver the selected power into a wide range of tissue types.
 
 
 

WEIGHT AND DIMENSIONS
HEIGHT: 11.1 CM (4 3/8")
WIDTH: 35.6 CM (14")
LENGTH: 43.9 CM (17")
WEIGHT: < 8.1 KG (< 18 LBS)
 
INPUT POWER REQUIREMENTS
 

Operating range is 85 to 132 ac volts. Maximum current is 7 amperes in cut and 4 amperes in coag.
 
 
 
HIGH FREQUENCY LEAKAGE
BIPOLAR: Less than 60 M ARMS
MONOPOLAR: Less than 150 M ARMS
 

REM™ CONTACT QUALITY MONITORING SYSTEM
MEASUREMENT FREQUENCY: 80 KHZ ± 10 KHZ
MEASUREMENT CURRENT: Less than  10 µA
ACCEPTABLE RESISTANCE RANGES:
        REM™ PAD - 5-135 OHMS
        NON-REM™ PAD - Less than  20 OHMS
Acceptance range is 5-135 OHMS after Rem Polyhesive™ II return electrode is applied.
 
ADAPTIVE REM™
Rem™ T rip is baseline impedance plus 40%. For example, if the baseline impedance is 30 ohms, the upper level trip is approximately 42 OHMS. If the pad-patient impedance falls below the baseline impedance, a new baseline is established.
 
 
OUTPUT CHARACTERISTICS
MODE MAXIMUM P-P VOLTAGE RATED LOAD (OHMS) MAXIMUM POWER (WATTS) CREST FACTOR (TYPICAL)
BIPOLAR PRECISE 450 100 70 1.5
STANDARD 320 100 70 1.5
MACRO 750 100 70 1.5
MONOPOLAR CUT LOW 1350 300 300 1.5
PURE 2300 300 300 1.5
BLEND 3300 300 200 2.5
MONOPOLAR COAG DESICCATE 3500 500 120 5.0
FULG. HIGH CRESTFACTOR 8500 500 120 7.0
FULG. LOW CREST FACTOR 6900 500 120 5.5
SPRAY 9000 500 120 8.0
CEM MODE MONOPOLAR CUR (LOW) 1000 300 100 1.5
MONOPOLAR COAG (DESICCATE) 3500 500 70 5.0
*Crest Factor ia an indicator of a waveform's ability to coagulate without cutting.

To contact us please call our toll-free number at 888-282-8644, local number at 407-322-3672 or fill out and submit this form.

* indicates required
Name: 
Address: 
City: 
State: 
Zip/Postal Code: 
Country: 
Phone: 
Fax: 
E-mail: 
Are you interested in receiving emails for upcoming product specials?:  Yes
This Is CAPTCHA Image
Write the characters in the image above
Your Comments and Questions: 
Blanket Warmers Surgical Lights Stretchers Sterilizers Scrub Sinks Surgical Tables Patient Monitors Anesthesia Machines Crash Carts / Anesthesia Carts
Future Health Concepts
1211 30th Street
Sanford, FL 32773
Email: Sales| Parts and Service
Phone: 1-888-282-8644
  1-407-322-3672
Fax: 1-407-322-3871
Home | Site Map | About Us | News | Contact Us

© 2008 by Future Health Concepts, All Rights Reserved.
Health Care Internet Marketing by SMT

Browse Medical Equipment

Anesthesia Machines | Blanket Warmers | Fluid Warmer | Medical Cart / Emergency Crash Carts
Cart Accessories | Defibrillators | Electrosurgical Units | Medical Gas Systems
Patient Monitor | Scrub Sinks | Stainless Steel Products | Stretchers
Sterilizers / Autoclaves | Surgical Lights | Surgical Tables | Surgical Table Accessories
Ultrasonic Cleaners | X-Ray Supplies and Equipment |

Medical Equipment Information

L&T Monitors | Operating Room Lights | Hospital Stretchers | Warming Cabinets | Hospital Equipment
Automated External Defibrillators | Anesthesia Equipment | Cardiac Defibrillator | Amsco Sterilizers
Electrosurgical Units | Instrument Washers