FHC-VALLEYLABEZ
FHC Valleylab EZ Electrosurgical Generator
OUTPUT WAVE FORMS:
BIPOLAR: 470 KHZ sinusoid
MONOPOLAR CUT:
PURE: 393 KHZ sinusoid
BLEND: 393 KHZ bursts of sinusoid, recurring at 27 KHZ intervals.
50% duty cycle envelope
MONOPOLAR COAG:
DESICCATION:
LOW 1: 240± 40 KHZ sinusoid recurring at 39 KHZ 8% duty cycle
LOW 2: 393 KHZ sinusoid
LOW 3: 393 KHZ sinusoid
FULGURATION:
HIGH 1: 470 ± 40 KHZ damped sinusoidal bursts with repetition
frequency of 57 KHZ
HIGH 2: 470 ± 40 KHZ damped sinusoidal bursts with repetition
frequency of 30 KHZ
Output power changes by less than 15% or 5 watts, whichever is
greater, as the line voltage varies from 90-135 volts and 186-264
volts (into a 300 OHM load)
LOW FREQUENCY LEAKAGE (50-60 HZ)
Source current, patient leads, all outputs tied together.
Normal polarity, intact chassis ground < 10 µA
Normal polarity, ground open< 50 µA
Reverse polarity, ground open< 50 µA
Sink current, 140V applied, all inputs < 50 µA
POWER READOUTS
Agree with actual power into rated load to within ± 15% or 5 watts, whichever is greater. Dosage error monitor.
PER = 97
Per (power efficiency rating) is the measure of an electrosurgical generator's ability to accurately deliver the selected power into a wide range of tissue types.
WEIGHT AND DIMENSIONS
HEIGHT: 12.7 CM (5")
WIDTH: 40.6 CM (16")
LENGTH: 39.5 CM (15.6")
WEIGHT: < 6.8 KG (< 15 LBS)
INPUT POWER REQUIREMENTS
110-120 VAC 220-240 VAC
Operating Range 58-140 VAC 170-280 VAC
Max Current Cut 8A 4A
Max Current Coag 4.2A 2.1A
HIGH FREQUENCY LEAKAGE
BIPOLAR: less than 60 M ARMS
MONOPOLAR: less than 150 M ARMS
REM™ CONTACT QUALITY MONITORING SYSTEM
MEASUREMENT FREQUENCY: 80 KHZ ± 10 KHZ
MEASUREMENT CURRENT: less than 10 µA
ACCEPTABLE RESISTANCE RANGES:
REM™ PAD - 5-135 OHMS
NON-REM™ PAD - less than 20 OHMS acceptance range is 5-135 OHMS after Rem Polyhesive™ II return electrode is applied.
ADAPTIVE REM™
Rem™ T rip is baseline impedance plus 40%. For example, if the
baseline impedance is 30 ohms, the upper level trip is
approximately 42 OHMS. If the pad-patient impedance falls below
the baseline impedance, a new baseline is established.
|
MODE |
MAXIMUM P-P VOLTAGE |
RATED LOAD (OHMS) |
MAXIMUM POWER (WATTS) |
CREST FACTOR (TYPICAL) |
|
|
|
|
|
|
| BIPOLAR |
STANDARD |
300 |
100 |
70 |
1.5 |
|
PURE |
2000 |
300 |
300 |
1.5 |
|
BLEND |
3400 |
300 |
200 |
2.1 |
| MONOPOLAR CUT |
DESICCATE LOW 1 |
3500 |
500 |
120 |
5.0 |
|
LOW 2 |
660 |
300 |
120 |
1.5 |
|
LOW 3 |
1100 |
300 |
120 |
1.5 |
|
FULGURATE HIGH 1 |
6200 |
500 |
120 |
5.0 |
|
HIGH 2 |
8500 |
500 |
120 |
7.0 |
*Crest Factor ia an indicator of a waveform's ability to coagulate without cutting.
|